ESR meters tend to use test signals of around 100 kHz and 100 mVAC. The low voltage prevents problems with applying reverse voltage to electrolytic capacitors and is low enough to keep semiconductors in associated circuits from effecting the results. At 100 kHz, ideal capacitors of 10uF or greater are essentially short circuits to the AC portion of the test signal. For such capacitors, the ESR can be found by either comparing the test results with those from resistors of similar value, or by calculating using the formula for parallel resistors.
A chart is provided at the bottom of this page to convert the voltage ratio to ESR. Capacitors down to about 1 uF will give meaningful results, but you will need to compare the signals to known good parts to interpret the signals
There are a couple of limitations to the simple version. If you were to accidentally test a capacitor with a few hundred volts of charge on it, the IC and some of the other parts would go up in smoke. The oscilloscope could also be damaged. The second limitation also applies to some commercial testers. Basically, a shorted capacitor and a low ESR capacitor test the same! Both of these limitations can be gotten around with a few modifications to the design.
If you test some capacitors in the .1 uF to 10 uF range at 100 kHz, you will note that as the value of the capacitor drops, the test signal distorts from a square wave. At 3.3 uF, the top and bottom of the wave form will have a distinct tilt, with a 1 uF capacitor, this effect is even more pronounced. The tilting effect is just capacitor charging up as it integrates the test signal. If you just want to check ESR, this is a nuisance, on the other hand if you want to know if the capacitor or associated circuitry is shorted..... a short does not integrate charge. By making the frequency variable, this effect can be used over a wide range of capacitors.
A second source of information is that with a DC coupled probe, the display will show a shift between a short
and a low ESR capacitor. A good capacitor under test will develop a DC+ bias at about 1/2 Vo which will shift the
waveform. A plain resistor or shorted component will not store charge, and so when the test signal
drops to zero, the signal on the oscilloscope will also show zero volts.
Other Additions
It's also possible to add protection to the tester to prevent inadvertent exposure to charged HV caps from damaging the tester and oscilloscope. Since the test voltage is so low, a couple pairs of back to back Si power diodes (D), combined with using power resistors for R5, R6, R7, and R8 should do the trick. It's probably a good idea to socket the IC anyway.
The third change to the second circuit is that the comparison resistor (R6) has a switchable parallel resistor (R7). The 2 ohm setting {switch closed) is better for checking low ESR caps, while the 10 ohm setting (switch open) is better for caps with higher ESR.
Using an AC millivoltmeter instead of an Oscilloscope
I tried the using a couple of DMM's to see if they could be used instead of an oscilloscope. In both cases, the results were quite usable with the following limitations. Both meters worked fine at 1 kHZ and 10 kHz, but at 100 kHz the results were dismal. Typical results (10 kHz, 100 mV P-P, with R6 = 5 ohms): Open circuit V= 38.4 mVAC, large cap with 2 ohm ESR V=10.8 mVAC, large cap with 5 ohm ESR V=18.8 mVAC.
These results are pretty much what one would expect from theory. The main limitation was operating frequency, at 10 kHz, the minimum capacitor that could be reasonably tested was 100 uF. If you have an old meter with a 1.5 volt range, it should be possible to redesign the circuit to make use of it. It would be a matter of using a 200ma min. power supply in the 12 - 15 volt range with R4 near the minimum value (75 ohm min@ 15 Volts to keep within the 555's current limit).
Additional comments:
The following table shows the ratio of peak to peak voltages for various values of ESR and R6-R7. Since R4 provides a current signal and V=IR, V is proportional to R. Using the ordinary formula for parallel resistors it is possible to construct a table. Note that R6 refers to the actual resistance of R6 and R7 in parallel.
ESR |
V/Vo |
V/Vo |
V/Vo |
(Ohm) |
(R6=2 Ohm) |
(R6=5 Ohm) |
(R6=10 Ohm) |
0.1 |
5% |
2% |
1% |
0.2 |
9% |
4% |
2% |
0.3 |
13% |
6% |
3% |
0.4 |
17% |
7% |
4% |
0.5 |
20% |
9% |
5% |
0.6 |
23% |
11% |
6% |
0.7 |
26% |
12% |
7% |
0.8 |
29% |
14% |
7% |
0.9 |
31% |
15% |
8% |
1 |
33% |
17% |
9% |
2 |
50% |
29% |
17% |
3 |
60% |
38% |
23% |
4 |
67% |
44% |
29% |
5 |
71% |
50% |
33% |
6 |
75% |
55% |
38% |
7 |
78% |
58% |
41% |
8 |
80% |
62% |
44% |
9 |
82% |
64% |
47% |
10 |
83% |
67% |
50% |
20 |
91% |
80% |
67% |
30 |
94% |
86% |
75% |
40 |
95% |
89% |
80% |
50 |
96% |
91% |
83% |
60 |
97% |
92% |
86% |
70 |
97% |
93% |
88% |
80 |
98% |
94% |
89% |
90 |
98% |
95% |
90% |
100 |
98% |
95% |
91% |
Copyright © 2000 by Stephen M. Powell
Comments on this page? E-mail smpowell@taxspam.usa.net
(remove the obvious extra eight character political statement from e-mail address)